DC Voltage range | 200 mV to 1000 V 5 ranges Best accuracy: ±0.010 % rdg. ±2 dgt. |
AC Voltage range | 2000 mV to 700 V 4 ranges Best accuracy: ±0.1 % rdg. ±100 dgt. Effective value rectifier |
Frequency characteristics at AC Voltage | 10 to 300 kHz |
Resistance range 2-terminal measurement | 200 ohm to 100 M-ohm 7 ranges Best accuracy: ±0.02 % rdg. ±2 dgt. |
Resistance range 4-terminal measurement | 200 ohm to 200 k-ohm 5 ranges Best accuracy: ±0.02 % rdg. ±2 dgt. |
DC Current range | 200 mA or 2000 mA 2 ranges Best accuracy: ±0.1 % rdg. ±6 dgt. |
AC Current range | 200 mA or 2000 mA 2 ranges Best accuracy: ±0.3 % rdg. ±100 dgt. (or used with clamp-on sensor range) |
Frequency characteristics at AC Current | 10 to 30 kHz (or clamp spec.) |
Frequency range | 100 Hz to 300 kHz 5 ranges Accuracy: ±0.015 % rdg. ±2 dgt. |
Continuity | Possible 50 ohm or less (at 2000 ohm range) |
Diode check | Possible (6 V open terminal voltage) |
Other functions | Current measurement (with clamp) Comparator external input and output Save/Load function AVE trigger RS-232C GP-IB ( -01 model ) |
Range switching | Auto or Manual |
Display | LED max. 199999 dgt. max. 999999 dgt. (at Frequency) |
Sampling rate | 300 times/sec (Fast) 8 to 9 time/sec (Medium) 1 time/sec (Slow) |
Power supply | 100 to 240 V AC 50/60 Hz (specify when ordering) |
Dimensions mass | 215 mm(8.46 in)W × 80 mm(3.15 in)H × 265 mm(10.43 in)D 2.6 kg (91.7 oz) |
Accessories | TEST LEAD 9170 (1) Power cord (1)
|